Why so many bed falls in care homes?
Within the next four weeks someone will be killed as a result of a bed fall.
Records indicate that 44,000 patients fell from their nursing beds in 2007 alone – of which eleven died. As the nighttime care assistant (charged with preventing a fall) is often faced with the prospect of dismissal, it is believed that many bed falls are not in fact reported.
It is also accepted that the cause of death as recorded by the GP can indicate cardiac arrest (natural causes) rather than a shock induced heart attack (which can result from a bed fall). If this is in fact true then the next of kin, the wider public and the British state are unaware as to the true extent of the tragedy.
Bed falls are not a new phenomena but their scale and frequency is. This is partially due to the ageing population, but also includes the drive to transfer tens of thousands of people from NHS open wards (c1998 – 2006) to traditional care homes and community care.
One major cause in the frequency of bed falls is that some care providers are unaware of their legal responsibilities and are providing the incorrect equipment.
Witten is at the heart of industrial Germany and is within easy commuting distance of several of Europe’s leading nursing bed manufacturers. In 1997, a team of experts met at the University of Witten, Nordrhein-Westfalen, Germany. Their purpose was to debate the issue of bed rail height with regards to a proposed safety standard for the design of an adjustable nursing bed and how to reduce both instances and the severity of bed falls.
When eventually the standard was released, it was entitled en1970 Adjustable Beds For Use By Disabled Persons and was adopted within the UK with the prefix BS.
BS en1970 focuses on the safe design and build of adjustable beds and compliance to this standard is a major part of the bed manufacturers certificate of European conformity otherwise known as a CE mark.
One problem that has arisen since the introduction of this critical safety standard is that some areas have been used out of context to encourage care providers to achieve a safe bed rail height. Whilst the motive has been to reduce bed falls, the consequence may have been to increase the severity and frequency of bed falls.
It is alleged that some local authorities have identified the safe height of bed rails as being 22cm (above the top of the uncompressed mattress) and are encouraging care providers to take measures to achieve this height – without providing the full background as to why this measurement is important and its relationship to other critical factors.
The height of bed rails at 22cm above the mattress was agreed upon at a time when most mattresses were of a foam construction and circa 12.5 cm (5”) in depth and so the bed rails were built proportionally to work with these mattresses.
In recent years, many of these static mattresses have been replaced with single layer dynamic air mattresses.
Dynamic air mattresses operate by sequentially deflating and inflating a series of air chambers during a given cycle. Whilst this has proven to be an effective aid in the treatment and prevention of pressure ulcers, they can sometimes trigger bed falls. This can be an acute problem in the community where the continuous monitoring of the patient is difficult. A single layer of deep cells can be unstable and render the bed rails ineffective.
Dynamic mattresses are often deeper than the benchmark 12.5cm either as a replacement or when used as an overlay mattress, and can therefore reduce the safe working height of the bed rail. This can invalidate the CE status of the bed, invalidate the insurance cover and challenge the legality of the care provider deploying the bed. The corporate manslaughter act of 2007 does not accept ignorance as an excuse for providing dangerous equipment.
It is estimated that in excess of 450,000 of these beds are currently in use.
Most care providers have simply used the correct dynamic hybrid mattresses of 12.5cm, and rest easy in terms of a bed rail height of 22cm.
Others have attempted to circumvent the issue through a profusion of adaptations, including Bed Rail Extensions, Underlays and Crash Mats.
Bed Rail Extensions
This goes completely against the original research into bed falls, as extensions (wooden or metal bars placed on top of the bed rail) create a higher barrier for the patient to climb over and fall from and hence create more serious injuries.
Other problems include:
• The dynamic mattress could in future be placed on a different make of bed frame without extensions or the extensions could simply be removed, thus exposing the original concerns. Most care providers offer multiple beds with various types of bed rails. The one sure option is to use a 12.5cm (5”) full replacement or dynamic hybrid.
• Deeper mattress combinations can be unstable which is a contributory factor in bed falls.
Crash Mats
Some care providers have decided to avoid the issue of bed rail heights by removing them completely. This is a great shame as bed rails, when used correctly, work and prevent thousands of injuries.
Crash Mats suggest that the patient will receive a soft landing, when in reality they may also be trip hazards – for both staff and patient – and can often have the effect of reducing impact in one area, only for the patient to hit their head against another surface, such as the bed or bedside cabinet.
Mattress Underlay
Overlay mattresses, as the name implies were created to overlay existing foam mattresses of about 12.5cm, thus providing sufficient pressure dissipation without bottoming out. Some care providers use the overlay without a mattress or with a foam base referred to as an underlay, of about 3cm depth.
This may change the pressure relieving qualities of the overlay mattress and expose the patient to pressure ulcers. It is possible to increase repositioning regimes but this substantial increase in cost is often higher than the cost of choosing the correct mattress.
Overlay dynamic mattresses are designed to sit on top of an existing mattress, whereas a 12.5cm dynamic hybrid will sit directly on the bed frame and provide the correct bed rail height.
Why do bed falls happen?
There are numerous cognitive, biological and behavioral reasons why a bed fall might occur and this can include anything from diet and urinary tract infections through to noise and disturbance.
The choice of mattress, where falls originate from, is a critical factor and the following notes detail why dynamics can effect bed falls and why modern hybrids of 12.5cm are an effective means of helping to prevent bed falls.
How do single layer dynamics work?
Dynamic air mattresses operate by sequentially deflating and inflating a series of air chambers during a given cycle. Whilst this methodology has proven to be an effective aid in the treatment and prevention of pressure ulcers they can sometimes trigger secondary issues such as bed falls. This can be an acute problem in the community where the continuous monitoring of the patient is difficult.
Disturbance
Some patients dislike the sensation of movement. This can be true for both overlays and replacements, and clients sometimes complain of a nauseous sensation similar to vertigo. Some patients find movement painful in terms of musculo skeletal injuries. Such disturbance can induce patients to attempt to get out of bed unassisted. Whilst this would stimulate effective intervention within most ward environments, in the community, constant monitoring is difficult, so falls are more likely.
Transfer Falls
Air cells sometimes deflate at the moment when a patient is preparing to stand or to transfer or whilst simply sitting on the edge of their bed. If the patient is attempting to support themselves on a deflating cell they can lose their balance and fall. Whilst dynamic mattress replacements may provide a static mode, the patient will sometimes attempt to sit, transfer or stand without assistance. Not all alternating mattresses provide a static mode, and they are often used where patients are expected to transfer without a hoist.
Further Resources
You can see more about pressure relief mattresses here
We also have an area dedicated to care beds and associated safety accessories